• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON BD MINIDRAW¿ FINGER SLEEVE, LARGE; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON BD MINIDRAW¿ FINGER SLEEVE, LARGE; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 366615
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/05/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E1: initial reporter addr 1: (b)(6).D2b: additional medical device type: gim, g5: additional pma / 510(k)#: k230391.The manufacturing location for this product is rmr.This site is an oem manufacturing site.Therefore, bd corporate headquarters in (b)(4) has been listed in sections d.3.And g.1.And the franklin lakes fda registration number has been used for the manufacture report number.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that when using one (1) bd minidraw¿ finger sleeve, large, there is blood smearing.No adverse impact was reported regarding the patient or user.
 
Additional Manufacturer Narrative
Investigation summary: bd did not receive samples for photos for investigation.Therefore, retention samples from bd inventory were evaluated by both visual examination and functional testing, and no issues were observed relating to blood pooling.The device was manufactured according to product specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode blood pooling.Bd was not able to identify a definitive root cause for the indicated failure mode.An update to the minidraw eifu was initiated to address the residual blood in finger sleeves that may occur.The new verbiage in the warning section will state: "practice universal precautions in accordance with your facility¿s procedures.Use gloves, gowns, eye protection, other personal protective equipment, and engineering controls to protect from residual blood on finger and/or finger sleeve, blood splatter, blood leakage, and potential exposure to blood-borne pathogens." additionally, the new text in step 8 of the blood collection procedure will state: "residual blood may be present on patient's finger and/or finger sleeve." complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.Bd was able to trace the root cause for the indicated failure mode to user technique.As part of bd's ongoing process to track and trend complaint data.
 
Event or Problem Description
It was reported that when using one (1) bd minidraw¿ finger sleeve, large, there is blood smearing.No adverse impact was reported regarding the patient or user.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MINIDRAW¿ FINGER SLEEVE, LARGE
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer Contact
matt j. connor
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key21529824
Report Number2243072-2025-00216
Device Sequence Number8339641
Product Code JKA
UDI-Device Identifier50382903666152
UDI-Public(01)50382903666152(11)240206(17)251202(10)3153001
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K230493
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date12/02/2025
Device Catalogue Number366615
Device Lot Number3153001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2025
Supplement Date Received by Manufacturer10/29/2025
Initial Report FDA Received Date03/05/2025
Supplement Report FDA Received Date10/30/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/06/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-