Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Correction: e4 h6: annex g code g04018 was inadvertently left out in initial emdr report.Additional information: d9 h3: device has been returned, and preliminary evaluation has been performed, however, our investigation is ongoing and device evaluation summary will be included in our follow up report once our investigation has been completed.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
It was reported, during a ureteral calculi stone removal procedure, the user detected the handle stuck and the tip damaged in a ngage nitinol stone extractor.So, its use was discontinued.The user changed to a new same type device to complete the procedure.No section of the device remained inside the patient's body.The patient did not require any additional procedures due to this occurrence.The complainant has not reported any adverse effects on the patient due to this occurrence.
|
|
Investigation evaluation: description of event: it was reported, during a ureteral calculi stone removal procedure, the user detected the handle stuck and the tip damaged in a ngage nitinol stone extractor.So, its use was discontinued.The user changed to a new same type device to complete the procedure.No section of the device remained inside the patient's body.The patient did not require any additional procedures due to this occurrence.The complainant has not reported any adverse effects on the patient due to this occurrence.Reviews of the complaint history, device history record (dhr), instructions for use (ifu), manufacturing instructions (mi), and quality control (qc) procedures, a visual inspection, and functional test of the returned device, were conducted during the investigation.One ngage nitinol stone extractor was returned in an open package with label.Handle in open position.Handle does not actuate basket formation.Support tubing loose from basket sheath.Evidence of adhesive present.Able to manually actuate basket formation.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record (dhr) found no non-conformances related to the reported failure mode.A review of complaint history records shows no other related complaints associated with device lot.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that that the device was manufactured to specification and there is no evidence that nonconforming product exists in house or in the field.A review of the ifu provides the following information: suggested handling instructions for extractors and forceps: important: excessive force could damage device.Cook has concluded that the cause for the complaint could not be established.The appropriate personnel have been notified, and cook will continue to monitor for similar events.Per the quality engineering risk assessment, no further action is required.Lastly, this report is being sent to indicate the complaint event is not reportable under fda 21 cfr part 803.An mdr guidance for manufacturers issued in 1997 stated that once a malfunction has caused or contributed to a death or serious injury, a presumption that the malfunction is likely to cause or contribute to a death or serious injury has been established.This presumption will continue until either the malfunction has caused or contributed to no further deaths or serious injuries for two years, or the manufacturer can show through valid data that the likelihood of another death or serious injury as a result of the malfunction is remote.A validated and detailed search of cook¿s complaint handling software revealed that there have been no deaths or serious injuries per 21 cfr part 803.3, due to the ngage nitinol stone extractor outer support sheath separating from the strain relief, from 01jan2022 through 11aug2025.Therefore, cook will cease malfunction reporting for events where the ngage nitinol stone extractor outer support sheath separated from the strain relief.The recurrence of a serious injury or death for the same malfunction will trigger the resumption of mandatory reporting, per 21 cfr part 803.50.As such, this event no longer meets the set criteria for a reportable event; no further reports regarding this event will be submitted.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|