• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC NGAGE NITINOL STONE EXTRACTOR; FFL DISLODGER, STONE, BASKET URETERAL, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC NGAGE NITINOL STONE EXTRACTOR; FFL DISLODGER, STONE, BASKET URETERAL, METAL Back to Search Results
Model Number G48294
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/24/2025
Type of Reportable Event Malfunction
Event or Problem Description
There is no new patient or event information to report.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Correction: e4 h6: annex g code g04018 was inadvertently left out in initial emdr report.Additional information: d9 h3: device has been returned, and preliminary evaluation has been performed, however, our investigation is ongoing and device evaluation summary will be included in our follow up report once our investigation has been completed.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.E1 - phone number: (b)(6).G4 - pma/510(k) #: exempt.H3 - device evaluation anticipated, but not yet begun.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
It was reported, during a ureteral calculi stone removal procedure, the user detected the handle stuck and the tip damaged in a ngage nitinol stone extractor.So, its use was discontinued.The user changed to a new same type device to complete the procedure.No section of the device remained inside the patient's body.The patient did not require any additional procedures due to this occurrence.The complainant has not reported any adverse effects on the patient due to this occurrence.
 
Additional Manufacturer Narrative
Investigation evaluation: description of event: it was reported, during a ureteral calculi stone removal procedure, the user detected the handle stuck and the tip damaged in a ngage nitinol stone extractor.So, its use was discontinued.The user changed to a new same type device to complete the procedure.No section of the device remained inside the patient's body.The patient did not require any additional procedures due to this occurrence.The complainant has not reported any adverse effects on the patient due to this occurrence.Reviews of the complaint history, device history record (dhr), instructions for use (ifu), manufacturing instructions (mi), and quality control (qc) procedures, a visual inspection, and functional test of the returned device, were conducted during the investigation.One ngage nitinol stone extractor was returned in an open package with label.Handle in open position.Handle does not actuate basket formation.Support tubing loose from basket sheath.Evidence of adhesive present.Able to manually actuate basket formation.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record (dhr) found no non-conformances related to the reported failure mode.A review of complaint history records shows no other related complaints associated with device lot.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that that the device was manufactured to specification and there is no evidence that nonconforming product exists in house or in the field.A review of the ifu provides the following information: suggested handling instructions for extractors and forceps: important: excessive force could damage device.Cook has concluded that the cause for the complaint could not be established.The appropriate personnel have been notified, and cook will continue to monitor for similar events.Per the quality engineering risk assessment, no further action is required.Lastly, this report is being sent to indicate the complaint event is not reportable under fda 21 cfr part 803.An mdr guidance for manufacturers issued in 1997 stated that once a malfunction has caused or contributed to a death or serious injury, a presumption that the malfunction is likely to cause or contribute to a death or serious injury has been established.This presumption will continue until either the malfunction has caused or contributed to no further deaths or serious injuries for two years, or the manufacturer can show through valid data that the likelihood of another death or serious injury as a result of the malfunction is remote.A validated and detailed search of cook¿s complaint handling software revealed that there have been no deaths or serious injuries per 21 cfr part 803.3, due to the ngage nitinol stone extractor outer support sheath separating from the strain relief, from 01jan2022 through 11aug2025.Therefore, cook will cease malfunction reporting for events where the ngage nitinol stone extractor outer support sheath separated from the strain relief.The recurrence of a serious injury or death for the same malfunction will trigger the resumption of mandatory reporting, per 21 cfr part 803.50.As such, this event no longer meets the set criteria for a reportable event; no further reports regarding this event will be submitted.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NGAGE NITINOL STONE EXTRACTOR
Common Device Name
FFL DISLODGER, STONE, BASKET URETERAL, METAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key21530011
Report Number1820334-2025-00221
Device Sequence Number9712602
Product Code FFL
UDI-Device Identifier00827002482944
UDI-Public(01)00827002482944(17)270722(10)16167019
Combination Product (Y/N)N
Initial Reporter CountryCN
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG48294
Device Catalogue NumberNGE-017115
Device Lot Number16167019
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/24/2025
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date03/05/2025
Supplement Report FDA Received Date03/27/2025
08/15/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/22/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-