MENTOR TEXAS MENTOR MEMORYGEL XTRA BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
SMHX500 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Post Operative Wound Infection (2446)
|
| Date of Event |
10/01/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made. as such, the investigation will be closed. if the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803. this report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date. this report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report. if certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: wound infection, rupture manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a breast augmentation revision surgery with implantation of a 500cc mentor memorygel xtra breast implant prosthesis.Post-operatively, the patient was diagnosed with a left breast infection and a ruptured left breast implant by a medical professional.As a result, the patient underwent unilateral left-sided breast implant removal without replacement on (b)(6) 2024.The patient is scheduled another surgery at a later date to implant a new left device and exchange the right device.No issues were reported for the right side.
|
| |
|
Additional Manufacturer Narrative
|
|
On march 10, 2025, the mentor analysis lab received the suspect medical device for evaluation.On march 17, 2025, mentor completed an evaluation on the returned device.Mentor then conducted a visual inspection, leak testing, and microscopic examination of the device.Visual inspection of the returned device determined that a tear (a) was observed on the anterior view of the breast implant, measuring approximately 0.6 cm.Leak testing was performed, according to the mentor procedure, and it revealed a tear (b) on the anterior view, measuring approximately 0.3 cm.No other tears were found during the analysis.Microscopic examination was performed on the edges of the ruptures (a & b), and parallel striations were found in the whole area of both tears.Parallel striations are consistent with markings made by a sharp object perforating the implant shell.Based on the information currently available, the microscopic examination of the returned product indicates that the implant could have been damaged during or subsequent to implantation.The product insert data sheet cautions to not allow cautery devices or sharp instruments, such as scalpels, suture needles, hypodermic needles, hemostats, adson forceps, or scissors to contact the device during the implantation or other surgical procedures.Infection, manifested by swelling, tenderness, pain, and fever, may appear in the immediate postoperative period or at any time after insertion of the implant.In the absence of classic symptoms, subacute or chronic infections may be difficult to diagnose.If infection does not subside promptly with the appropriate treatment, removal of the implant is indicated.Infection is a known complication associated with any surgery.Per a database query, this is the only reported complaint of infection against this lot number.The mentor microbiology department has provided information on the sterility lot for the implanted device indicating that it met all sterilization parameters required to provide a 10e-6 sterility assurance level prior to release for distribution.Product evaluation concluded the reported infection occurred from a source other than the device.Infection is a known complication associated with any surgery and is referenced in our current product insert data sheet.Manufacturer¿s reference number: (b)(4).
|
| |
|