| Catalog Number |
0250070600 |
| Medical Device Problem Code |
Device Contaminated During Manufacture or Shipping (2969)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/05/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that foreign material found inside the sterile packaging.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: per rep, "i have an account that found a blue item inside their sterile ahto packaging when opening for a case today." the failure(s) identified in the investigation is consistent with the complaint record.The probable root causes could be the tyvek cover was dirty or the packagers hand had contaminants during packaging.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that foreign material found inside the sterile packaging.
|
| |
|