• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PFC SIGMA TIB TRAY CEM SZ3; SIGMA KNEE PRIMARY : KNEE TIBIAL TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PFC SIGMA TIB TRAY CEM SZ3; SIGMA KNEE PRIMARY : KNEE TIBIAL TRAY Back to Search Results
Catalog Number 866023
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Codes Pocket Erosion (2013); Unspecified Tissue Injury (4559); Bone Notching (4890)
Date of Event 11/06/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that patient underwent revision surgery, requiring extensive debridement.Extensive bone erosion was noted on the medial and lateral femoral condyles as well as along the medial tibial plateau requiring bone grafting to correct.Granulation tissue indicative of loosening was noted.The femoral and tibial sides were noted to be loose - interface not identified.The patella was removed and underneath small cysts were identified and removed before replacing.A fracture of the femoral stem and the connecting screw occurred without trauma.The connecting screw between the femoral shield and the femoral shaft is affected.The revision went smoothly.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: d6, h4: the information provided regarding the manufacturing date and implantation date are identified to be conflicting.Continued follow-up is being conducted and any additional information received will be reported.
 
Additional Manufacturer Narrative
Product complaint#: (b)(4).Investigation summary: user report received bfarm case: (b)(4).In the above-mentioned partially coupled, cemented revision prosthesis, a fracture of the femoral stem and the connecting screw occurred without trauma.Femoral stem and the connecting screw in the connection area to the femoral shield the product was not returned to j&j medtech orthopaedics; however, photos were provided for review.The photo investigation revealed that the pfc sigma tib tray cem sz3 had organic material on the surface.No cement is seen on the visible fixation surfaces of the tibial tray.However, the photos provided do not display the entire fixation surface, and the sections that are visible do not exhibit any indications of burnishing that might suggest micromotion.Therefore, implant loosening is not confirmed.A manufacturing record evaluation was performed for the finished device [866023 / 2582554] number, and no non-conformance's / manufacturing irregularities related to the malfunction were identified.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was not confirmed as the observed condition of the pfc sigma tib tray cem sz3 would not have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: 1) quantity manufactured: (b)(4).2) date of manufacture: 3/12/2008.3) any anomalies or deviations identified in dhr: none.4) expiry date: 3/20/2018.5) ifu reference: ifu-0902-00-252.Device history review: a manufacturing record evaluation was performed for the finished device [866023 / 2582554] number, and no non-conformance's / manufacturing irregularities related to the malfunction were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFC SIGMA TIB TRAY CEM SZ3
Common Device Name
SIGMA KNEE PRIMARY : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key21536909
Report Number1818910-2025-03544
Device Sequence Number20767243
Product Code JWH
UDI-Device Identifier10603295216780
UDI-Public(01)10603295216780(17)20180320(10)2582554
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K884796
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/20/2018
Device Catalogue Number866023
Device Lot Number2582554
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/14/2025
Supplement Date Received by Manufacturer06/18/2025
Initial Report FDA Received Date03/06/2025
Supplement Report FDA Received Date06/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/20/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-