This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: d6, h4: the information provided regarding the manufacturing date and implantation date are identified to be conflicting.Continued follow-up is being conducted and any additional information received will be reported.
|
|
Product complaint#: (b)(4).Investigation summary: user report received bfarm case: (b)(4).In the above-mentioned partially coupled, cemented revision prosthesis, a fracture of the femoral stem and the connecting screw occurred without trauma.Femoral stem and the connecting screw in the connection area to the femoral shield the product was not returned to j&j medtech orthopaedics; however, photos were provided for review.The photo investigation revealed that the pfc sigma tib tray cem sz3 had organic material on the surface.No cement is seen on the visible fixation surfaces of the tibial tray.However, the photos provided do not display the entire fixation surface, and the sections that are visible do not exhibit any indications of burnishing that might suggest micromotion.Therefore, implant loosening is not confirmed.A manufacturing record evaluation was performed for the finished device [866023 / 2582554] number, and no non-conformance's / manufacturing irregularities related to the malfunction were identified.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was not confirmed as the observed condition of the pfc sigma tib tray cem sz3 would not have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: 1) quantity manufactured: (b)(4).2) date of manufacture: 3/12/2008.3) any anomalies or deviations identified in dhr: none.4) expiry date: 3/20/2018.5) ifu reference: ifu-0902-00-252.Device history review: a manufacturing record evaluation was performed for the finished device [866023 / 2582554] number, and no non-conformance's / manufacturing irregularities related to the malfunction were identified.
|