This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Product complaint # (b)(4).Investigation summary : user report received bfarm case (b)(4).In the above-mentioned partially coupled, cemented revision prosthesis, a fracture of the femoral stem and the connecting screw occurred without trauma.Femoral stem and the connecting screw in the connection area to the femoral shield.The product was not returned to j&j medtech orthopaedics, however photos were provided for review.See attachment (b)(4), 20080411 used implant, 20080411 photos explants.The photo investigation revealed that the device had organic material on the surface.It is worth noting that there is an excessive cement on the device.Cement appears to be applied in excess.Although a definitive cause for device fracture is not established, one contributing factor appears to be loosening at the interface between the cement and bone.The specific manufacturer of the cement remains unknown; however, it is evident that an excess amount of cement was used to fill bone voids in the anterior portion of the distal femoral head.This likely resulted in inadequate fixation between the cement and the bone.Insufficient fixation can lead to micromotion, which may increase stress at the junction of the stem bolt and the stem.This can result in fatigue cracks and ultimately catastrophic failure.A manufacturing record evaluation was performed for the finished device [960088 / 336088ad], and no non-conformances or manufacturing irregularities were identified.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was not confirmed as the observed condition of the pfc*sigma tc3 fem rt sz3 would not have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : 1) quantity manufactured: 6.2) date of manufacture: 2006-08.3) any anomalies or deviations identified in dhr: none.4) expiry date: 2016-08.5) ifu reference: ifu-0902-00-252.Device history review : a manufacturing record evaluation was performed for the finished device [960088 / 336088ad], and no non-conformances or manufacturing irregularities were identified.Added: h4.Corrected: d4 (expiry date, primary udi number), h6.
|