| Catalog Number |
960866 |
| Medical Device Problem Code |
Loss of or Failure to Bond (1068)
|
| Health Effect - Clinical Codes |
Pocket Erosion (2013); Unspecified Tissue Injury (4559); Bone Notching (4890)
|
| Date of Event |
11/06/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that patient underwent revision surgery, requiring extensive debridement.Extensive bone erosion was noted on the medial and lateral femoral condyles as well as along the medial tibial plateau requiring bone grafting to correct.Granulation tissue indicative of loosening was noted.The femoral and tibial sides were noted to be loose - interface not identified.The patella was removed and underneath small cysts were identified and removed before replacing.A fracture of the femoral stem and the connecting screw occurred without trauma.The connecting screw between the femoral shield and the femoral shaft is affected.The revision went smoothly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: d6, h4: the information provided regarding the manufacturing date and implantation date are identified to be conflicting.Continued follow-up is being conducted and any additional information received will be reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: user report received bfarm case (b)(4).In the above-mentioned partially coupled, cemented revision prosthesis, a fracture of the femoral stem and the connecting screw occurred without trauma.Femoral stem and the connecting screw in the connection area to the femoral shield the product was not returned to j&j medtech orthopaedics, however photos were provided for review.The photo investigation revealed nothing indicative of a device nonconformance.There is no evidence that suggests that a loosening condition occurred between the implant-cement interface.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was not confirmed as the observed condition of the pfc sigma fem post aug sz3 4mm would not have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a search of the legacy non-conformance quality system found no nonconformances against this part and lot.Corrected: h4, h6.
|
| |
|
Search Alerts/Recalls
|