| Catalog Number |
XCR120306 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported by the field, during a coil embolization, the hypo tube of a freeform xsft 3mm x 6 cm coil (xcr120306, 31445669) slipped off the back when manual detachment happened.Coil still detached.There was no patient injury reported.Additional event information received on 04-mar-2025 indicating that the unruptured aneurysm was at middle cerebral artery, m1 segment.The device was not inadvertently subjected to force during the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Section d2b ¿ procode: krd/hcg.Section d4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The company is seeking this information through the event investigation.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's ref.No(b)(4).Updated sections on this medwatch: b4, b5, d4, g3, g6, h2, h4, h6 and h11.Section b5: based on additional event information received on 11-mar-2025, the correct lot number for the device involved is 31345669.Complaint conclusion: as reported by the field, during a coil embolization, the hypo tube of a freeform xsft 3mm x 6 cm coil (xcr120306, 31345669) slipped off the back when manual detachment happened.Coil still detached.There was no patient injury reported.Additional event information received on 04-mar-2025 indicating that the unruptured aneurysm was at middle cerebral artery, m1 segment.The device was not inadvertently subjected to force during the procedure.Product analysis cannot be conducted as the product was not returned for analysis.No determination of causes and possible contributing factors could be made.A manufacturing record evaluation was performed for the finished device 31345669 number, and no non-conformances related to the malfunction were identified.With the information available and without the product available for analysis, the reported customer complaint could not be confirmed.Based on the manufacturing record evaluation, there is no indication that the event is related to the device manufacturing process.The exact cause of the event could not be determined; however, there are circumstances of the procedure that may have contributed to the reported failure.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
|
| |
|
Search Alerts/Recalls
|
|