EXACTECH, INC. LOGIC CR FEMORAL POR, LEFT, SZ 3; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-010-04-0230
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/29/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(d10) concomitant devices: 02-012-49-3009 - logic cr tib insert slope++, sz 3, 9mm: (b)(6); 02-010-04-0230 - logic cr femoral por, left, sz 3: (b)(6); 200-02-35 - three peg patella 35mm: (b)(6); 201-78-15 - holding pin mini sharp point 4 pk: (b)(6); 201-78-81 - 3 trocar, mod.Hex 2pk: (b)(6); 521-78-23 - threaded pin size 2.3 collared: (b)(6); 521-78-23 - threaded pin size 2.3 collared: (b)(6); 521-78-32 - threaded pin size 3.0 collarless: (b)(6); a10012 - gps implant kit v2: (b)(6).The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Event or Problem Description
As reported, approximately four years post initial left side tka, the female patient had a revision due to poly wear.Patient was revised into a logic ps tkr.There was no breakage of device or surgical delay/prolongation.Patient was last known to be in stable condition following the event.Images received.Sales rep was unable to obtain x-rays.The devices are not available for evaluation as they were disposed at the hospital, no other patient information/medical history reported.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were corrected: d1, h6.The revision reported was likely the result of improper implant position, the overall posterior slope of the insert and/or potential ligamentous/soft tissue imbalances which allowed for increased posterior contact between the femoral component and the tibial insert and led to the observed damage of the polyethylene.Additionally, a contributing factor to the observed damage may have been the result of being packaged in a non-conforming vacuum bag for more than five years.However, this cannot be confirmed as the device was not returned for evaluation and radiographs were not provided.The most probable root cause associated with the reported event of ¿prosthesis wear¿ is associated with material damage to a surface, usually involving progressive loss or displacement of material, due to relative motion between that surface and a contacting substance or substances.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC CR FEMORAL POR, LEFT, SZ 3
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
MDR Report Key21538475
Report Number1038671-2025-01460
Device Sequence Number17917039
Product Code JWH
UDI-Device Identifier10885862158246
UDI-Public10885862158246
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K123687
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/02/2023
Device Catalogue Number02-010-04-0230
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2025
Supplement Date Received by Manufacturer03/25/2025
Initial Report FDA Received Date03/06/2025
Supplement Report FDA Received Date03/26/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/03/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;