• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. LOGIC FEMORAL PS CEM RIGHT SZ 3.5; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. LOGIC FEMORAL PS CEM RIGHT SZ 3.5; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number LOGIC FEMORAL PS CEM RIGHT SZ 3.5
Medical Device Problem Codes Unstable (1667); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Pain (1994); Joint Laxity (4526); Swelling/ Edema (4577)
Date of Event 08/22/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional, material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.D10 concomitant devices: (b)(6), 02-012-35-3509 - logic tibia ps mod insrt sz 3.5 9mm; (b)(6), 02-012-45-3535 - lgc tibial fit tray cem sz 3.5f / 3.5t; (b)(6), 200-02-35 - three peg patella 35mm.
 
Event or Problem Description
It was reported via legal documentation that approximately 117 months after a right total knee replacement procedure, the patient underwent a revision procedure to address polyethylene wear, pain, swelling, and instability.The revising surgeon noted loosening of the femoral component, as well as bony defects in both the femur and tibia.No further issues or complications were reported.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC FEMORAL PS CEM RIGHT SZ 3.5
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
MDR Report Key21539125
Report Number1038671-2025-01462
Device Sequence Number15387359
Product Code JWH
UDI-Device Identifier10885862001245
UDI-Public10885862001245
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K093360
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/17/2018
Device Model NumberLOGIC FEMORAL PS CEM RIGHT SZ 3.5
Device Catalogue Number02-010-01-0335
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date04/14/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/14/2024
Initial Report FDA Received Date03/06/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age67 YR
Patient SexFemale
-
-