Catalog Number 021.2114
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/31/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The batch number could be verified.Our manufacturing q-system assures, that production and process controls are in place to ensure that batches confirm to the applicable specifications before they are distributed.The fracture of a dental implant is a known long-term complication of endosseous dental implants.The manufacturers trend analysis confirms that the reported implant fracture rate associated with its dental implants is low and remains consistent with the experience as documented in the literature.There are different reasons why implants break and this problem has been illuminated by various authors (e.G.Maeglin 1988, beumer 1989, tetsch 1991, worthington 1992).Beumer and lewis 1989 report the following causes: a) production / design flaws b) inadequate fitting of the suprastructure c) occlusal overload d) bruxism e) bone resorption around the implants f) insufficient axle alignment of the implants g) trauma h) biomechanical causes i) design of the suprastructure.
 
Event or Problem Description
The clinician reports the implant was inserted (b)(6) 2017 in fdi 33.On (b)(6) 2025, fracture of the implant was verified.The device was forwarded to the manufacturer.There were no patient operative or post-operative complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BL, Ø 3.3 MM NC, SLACTIVE 14 MM
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
INSTITUT STRAUMANN AG
peter merian-weg 12
basel 04002
SZ  04002
Manufacturer (Section G)
INSTITUT STRAUMANN AG
peter merian-weg 12
basel 04002
SZ   04002
Manufacturer Contact
jennifer jackson
60 minuteman road
andover, MA 01810
9787472509
MDR Report Key21539234
Report Number0009613348-2025-002921
Device Sequence Number7071336
Product Code DZE
UDI-Device Identifier07630031706379
UDI-Public07630031706379
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K062129
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 03/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/14/2018
Device Catalogue Number021.2114
Device Lot NumberGH439
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/10/2025
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date03/06/2025
Event Location Other
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date03/06/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/14/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-