| Catalog Number |
0070310 |
| Medical Device Problem Code |
Decrease in Suction (1146)
|
| Health Effect - Clinical Code |
Hematoma (1884)
|
| Date of Event |
02/18/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the incident report was filed on as they had to return to surgery for a hematoma evacuation because the drain did not work as it should have.This is not the first patient that they had this problem with.This complaint is for product (pcn #0070310) and (pcn #0070740).
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed cause unknown.No actions can be taken at this time since a root cause was not identified.Photo: received (2) photo samples.All photo samples showcase an overview of a wound drain and tubing.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and state the following: indications: wound drains are used to remove exudates from wound sites.Warning: when placing drain(s) care should be taken to ensure that the perforated portion of the wound drain lies completely within the confines of the wound.Read product insert provided with the closed wound evacuator for detailed instructions, warnings and precautions associated with the use of the evacuator device.To avoid the possibility of drain damage or breakage: additional perforations should not be made in the drains.Avoid suturing through drains.Drains should lie flat and in line with the skin exit areas.Particular care should be taken to avoid any obstacles to the drain exit path.Drains should be checked for free motion during closure to minimize the possibility of breakage.Drain removal should be done gently by hand.They should not be handled with pointed, toothed or sharp instruments which could cause cuts or nicks and lead to subsequent structural failure of the drain.Surgical removal may be necessary if drain is difficult to remove or breaks.Drain placement: 1.Place perforated wound drain within the critical fluid collection area of wound.2.Draw non-perforated section of wound drain through to the outside until drain indicator mark appears at the skin surface.Two sets of indicator marks aid placement of the drain.3.Remove trocar only by cutting the drain one inch from end of trocar.4.Trim non-perforated section of drain to desired length.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the incident report was filed on as they had to return to surgery for a hematoma evacuation because the drain did not work as it should have.This is not the first patient that they had this problem with.This complaint is for product (pcn #(b)(4)) and (pcn #(b)(4)).
|
| |
|
Search Alerts/Recalls
|