ARROW INTERNATIONAL LLC ARROW CVC SET: 2-LUMEN 7 FR X 20 CM; CATHETER INTRAVASCULAR THERAPE Back to Search Results
Catalog Number CV-47702-E
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/25/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "the doctor found the swg kinked during used on the patient.".They changed a new one to resolve the issue.No harm for the patient.The patient condition is fine.No medical intervention was required.
 
Additional Manufacturer Narrative
(b)(4).
 
Additional Manufacturer Narrative
(b)(4).The customer returned one, opened cvc set including the guide wire within its advancer, an 18ga introducer needle, and an arrow raulerson syringe (ars) for analysis.Signs of use were observed on the returned components.The guide wire was observed to have one kink towards the distal j-bend.The distal j-bend was slightly misshapen but intact.Microscopic examination confirmed the kink in the guide wire body.Both welds were present and were observed to be full and spherical.The kink in the guide wire measured 21 mm from the distal tip.The overall length of the guide wire measured 601 mm, which is within the specification limits of 596-604 mm per guide wire product drawing.The outer diameter of the guide wire measured 0.814 mm, which is within the specification limits of 0.788-0.826 mm per guide wire product drawing.The guide wire was functionally tested per the product instructions for use (ifu).The ifu provided with this kit instructs the user, "advance guidewire into arrow raulerson syringe approximately 10 cm until it passes through syringe valves or into introducer needle." the guide wire was advanced through the returned 18ga introducer needle and ars to functionally test the guide wire.The undamaged portions of the guide wire passed through with little to no resistance.A manual tug test confirmed that both the distal and proximal welds were intact.A device history record review was performed, and no relevant findings were identified.The ifu provided with the kit describes suggested techniques to minimize the likelihood of guide wire damage during use.The instructions caution, "do not apply excessive force in placing or removing catheter or guidewire.Excessive force can cause component damage or breakage.If damage is suspected or withdrawal cannot be easily accomplished, radiographic visualization should be obtained and further consultation requested." the report that the guide wire kinked during use was confirmed through complaint investigation of the returned sample.The guide wire had one kink towards the distal j-bend.The returned guide wire met all relevant dimensional/functional requirements, and a device history record review did not identify any manufacturing related issues.Based on the condition of the guide wire and the report that the damage was observed during use, unintentional use error caused or contributed to this event.Teleflex will continue to monitor and trend for reports of this nature.
 
Event or Problem Description
It was reported that "the doctor found the swg kinked during used on the patient.".They changed a new one to resolve the issue.No harm for the patient.The patient condition is fine.No medical intervention was required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW CVC SET: 2-LUMEN 7 FR X 20 CM
Common Device Name
CATHETER INTRAVASCULAR THERAPE
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key21547889
Report Number3006425876-2025-00239
Device Sequence Number18155233
Product Code FOZ
UDI-Device Identifier10801902102536
UDI-Public(01)10801902102536(11)230419(17)250302(10)71F23D0911(21)
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K993691
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/02/2025
Device Catalogue NumberCV-47702-E
Device Lot Number71F23D0911
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/03/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/19/2025
Supplement Date Received by Manufacturer03/19/2025
Initial Report FDA Received Date03/07/2025
Supplement Report FDA Received Date03/31/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/19/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED; NOT REPORTED
Patient SexUnknown
-
-