|
SMITH & NEPHEW, INC. FLEXIBLE SHAFT W/CIR CONNECTOR; PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED
|
Back to Search Results |
|
| Catalog Number |
71118200 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during an internal fixation surgery, when the flexible shaft w/cir connector was removed it was found to have fractured in its distal portion, but without being disassembled from the device.The procedure was resumed, without any delay, using the same device.Patient was not harmed as consequence of this problem.The patient's health condition is stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during an internal fixation surgery, when the flexible shaft w/cir connector was removed it was found to have fractured in its distal portion, but without being disassembled from the device, then no piece fell into the patient.The procedure was resumed, without any delay, using the same device.Patient was not harmed as consequence of this problem.The patient's health condition is stable.
|
| |
|
Search Alerts/Recalls
|
|
|