• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG ILESTO 7 VR-T DX DF-1; ICD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG ILESTO 7 VR-T DX DF-1; ICD Back to Search Results
Model Number 383594
Medical Device Problem Code Premature End-of-Life Indicator (1480)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/08/2025
Type of Reportable Event Malfunction
Event or Problem Description
This device was explanted due to eos.No adverse patient events were reported.Should additional information become available, this file will be updated.
 
Additional Manufacturer Narrative
Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this icd were reviewed.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.The icd interrogation revealed the eos battery status, detected on december 14, 2024.The memory content of the icd was inspected, revealing a normal current consumption of this device.However, the amount of charge taken from the battery was verified and the battery condition was not as expected.Therefore, the icd was opened and the inner assembly was inspected.The visual inspection of the inner assembly showed no anomalies.In a next step, the electronic module was attached to an external power supply and the eos status was removed with a technical programmer to check the functionality of the electronic module.The overall measured current consumption was normal and as expected.Analysis of the electronic module revealed no indication of a malfunction.Therefore, the battery was sent to the manufacturer for further analysis.The manufacturing records of the battery were inspected, documenting that the battery parameters were within specification during the battery manufacturing.No anomalies were documented during the production process associated with this battery.The visual inspection of the battery did not reveal any external signs of damage.In a next step, the battery was opened for destructive analysis.During the analysis of the inner assembly an elevated internal battery impedance was identified.It is reasonable to assume that the increased impedance has led to a voltage drop and therefore to the clinical observation.In conclusion, the electronic module of the icd was proven to be fully functional.Further analysis revealed an elevated battery impedance as the root cause of the clinical observation.
 
Event or Problem Description
This device was explanted due to eos.No adverse patient events were reported.Should additional information become available, this file will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ILESTO 7 VR-T DX DF-1
Common Device Name
ICD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
DE  12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key21549351
Report Number1028232-2025-01195
Device Sequence Number19946024
Product Code LWS
UDI-Device Identifier04035479125509
UDI-Public(01)04035479125509(17)160229
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P050023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 09/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2016
Device Model Number383594
Device Catalogue Number383594
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2025
Supplement Date Received by Manufacturer09/08/2025
Initial Report FDA Received Date03/07/2025
Supplement Report FDA Received Date09/08/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/06/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age83 YR
Patient SexMale
-
-