| Catalog Number |
A0756225G0M03 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Summary: returned waveone gold medium file 25mm is actually broken at the tip of the active part (torque).No material defect was found during analysis of the rupture pattern.No unused file is available for evaluation.Nothing unusual to report was found during dhr review (batch # 1891155).Root causes are not identified.We will track this kind of event and monitor the trend.
|
| |
|
Event or Problem Description
|
|
In this event it is reported that a waveone gold medium 6-file ster 25 mm broke during use.The broken part remains in the canal after attempts were made.It may be necessary to perform apical surgery or extraction of tooth.
|
| |
|
Additional Manufacturer Narrative
|
|
As a result of this malfunction, the potential for surgical intervention exists to preclude permanent damage to a body structure or permanent impairment of a body function as evidenced by previous reported events with similar files.This event, therefore, is reportable per 21cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|