| Model Number |
439688 |
| Medical Device Problem Codes |
Failure to Capture (1081); Adverse Event Without Identified Device or Use Problem (2993); High Capture Threshold (3266)
|
| Health Effect - Clinical Code |
Undesired Nerve Stimulation (1980)
|
| Date of Event |
01/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient experienced phrenic nerve stimulation from the left ventricular (lv) lead.It was also noted that the lv lead exhibited high thresholds and occasional loss of capture.The lead remains in use.No further patient complications have been reported as a result of this event.
|
| |
|
Search Alerts/Recalls
|