| Catalog Number |
520001040 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Foreign Body Reaction (1868); Pain (1994)
|
| Date of Event |
02/14/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient came in complaining of pain a 1-2 weeks ago.A ct scan was done and it was discovered that the inhance reverse liner could be seen floating in the shoulder and was not attached to the humeral shell/stem construct.It also showed what appeared to be metallosis indicating that the glenosphere had been rubbing on the face of the humeral shell.Revision surgery was done on (b)(6) 2025.There was a significant amount of metal debris and metallosis on the bone and the surrounding tissue.It is unknown how long the poly had been out of place but by the damage done to the implants as well as the amount of metallosis, it seemed to have happened quite a while back.The glenosphere had worn a divot in the face of the shell.Surgeon was able to remove the glenosphere, the humeral shell, the stem, and retrieved the poly.The baseplate and screws were well fixed and stayed implanted.The surgeon cleaned/removed the metal debris and metallosis out of the bone and tissue.The surgeon was able to revise to a larger and longer stem and achieve good fixation, he put on a new glenosphere of the same size and trialed a few humeral shell and poly combinations to determine what was best.When x-raying to check our positioning of new parts, an acromial fracture could be seen.The preferred combination of a +8 humeral shell for an xl stem and a 40 +0 retentive liner was chosen and implanted by the surgeon.Doi: unknown.Affected side: right shoulder.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.A manufacturing record evaluation (nc search) was performed for the finished device product code: 520001040, lot -659353 and no non-conformances / manufacturing irregularities were identified.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot a manufacturing record evaluation (nc search) was performed for the finished device product code: 520001040, lot -659353 and no non-conformances / manufacturing irregularities were identified.
|
| |
|
Search Alerts/Recalls
|