|
EXACTECH, INC. TRULIANT PS CEM FEM PS CEM LEFT SZ 4; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-020-11-0240 |
| Medical Device Problem Codes |
Unstable (1667); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Pain (1994); Scar Tissue (2060); Synovitis (2094); Osteolysis (2377); Joint Laxity (4526); Swelling/ Edema (4577)
|
| Date of Event |
01/29/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that approximately 75 months after a left total knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear, premature polyethylene wear from recalled polyethylene insert, aseptic loosening of femoral and patellar components.Revision surgery operative notes were provided.Patient left the operating room in stable condition.Post operative diagnosis noted wear of the implant.No further issues or complications were reported.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
D10: (b)(6) - 02-022-35-4015 - truliant tib imp ps insert sz 4 15mm.(b)(6) - 02-022-45-4040 - truliant tib fit tray cem sz 4f / 4t.(b)(6) - 02-012-60-1425 - tru stem ext 14mm x 25mm.(b)(6) - 02-020-11-0240 - truliant ps cem fem ps cem left sz 4.(b)(6) - 204-70-00 - tibial stem ext.Screw.(b)(6) - 200-07-32 - advanced patella 32mm 3 peg implant.Multiple mdr reports were filed for this event, please see associated reports: 1038671-2024-01315, 1038671-2025-01496.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, instability, and patellar loosening and femoral loosening or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|