• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE "PLASMALOOP", LOOP, MEDIUM, 24 FR., 12°-30°, ESG TURIS; WA22706S

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE "PLASMALOOP", LOOP, MEDIUM, 24 FR., 12°-30°, ESG TURIS; WA22706S Back to Search Results
Model Number WA22706S
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/08/2024
Type of Reportable Event Malfunction
Event or Problem Description
During the device evaluation, the resection electrode exhibited the cutting loop was bent down and had charred marks.There were no reports of patient involvement.
 
Additional Manufacturer Narrative
Based on the results of the investigation, it is likely that the following led to the malfunction: excessive force.A definitive root cause cannot be identified.A device history review revealed no issues that could have caused or contributed to the reported issue.Olympus will continue to monitor the field performance of this device.Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
Correction is being made to previously submitted g3 - date received by manufacturer.The correct date was 05sep2024.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE "PLASMALOOP", LOOP, MEDIUM, 24 FR., 12°-30°, ESG TURIS
Common Device Name
WA22706S
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key21554217
Report Number9610773-2025-80014
Device Sequence Number8165234
Product Code FAS
UDI-Device Identifier14042761085356
UDI-Public14042761085356
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K171965
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberWA22706S
Device Lot Number1000130675
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/17/2025
Supplement Date Received by Manufacturer04/10/2025
Initial Report FDA Received Date03/07/2025
Supplement Report FDA Received Date04/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/13/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-