| Model Number |
WA22706S |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/08/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During the device evaluation, the resection electrode exhibited the cutting loop was bent down and had charred marks.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the results of the investigation, it is likely that the following led to the malfunction: excessive force.A definitive root cause cannot be identified.A device history review revealed no issues that could have caused or contributed to the reported issue.Olympus will continue to monitor the field performance of this device.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction is being made to previously submitted g3 - date received by manufacturer.The correct date was 05sep2024.
|
| |
|