| Model Number |
DSX1030S11 |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging visualization of particles in device.Related to a cpap device's sound abatement foam.There was no report of patient harm or injury.There was no report of medical intervention.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging visualization of particles in device.Related to a cpap device's sound abatement foam.There was no report of patient harm or injury.The device was returned to the manufacturer service center for further evaluation.During the evaluation of the device at the manufacturer service center, the device was visually inspected and found no evidence of foam degradation.The device powered on and airflow was confirmed.The device's downloaded logs were reviewed by the manufacturer service center.There was 0 error code found.The manufacturer service center concludes that they couldn't confirm the customer's allegation and there was no visible foam degradation.Box h6: evaluation method code grid, evaluation results code and conclusion code grid has been updated.
|
| |
|
Search Alerts/Recalls
|