Additional information: d9, h3, h4, h6, h11.H4: the lot was manufactured between june 28-29, 2023.H11: two devices (1 opened and used sample and 1 unopened companion sample) were received for evaluation.Visual inspection was performed on all the samples, and the reported issue of leakage was not easily identified on any of the returned samples.Functional testing was performed with a main module compounding system performing all set up and actual compounding, and the leak(s) were not replicated.The products operated as intended and no issues or defects were found or encountered on the actual and the companion samples.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|