• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. XPS® SYSTEM NAVIGATED BLADE; BUR, EAR, NOSE AND THROAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. XPS® SYSTEM NAVIGATED BLADE; BUR, EAR, NOSE AND THROAT Back to Search Results
Model Number 1884380EM
Medical Device Problem Codes Break (1069); Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during procedure, shaver the blade got stuck and was broken in shaft.There was no patient impact.
 
Additional Manufacturer Narrative
H3: analysis results were not available as of the date of this report.A follow-up report will be submitted when analysis is complete.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
H3: product analysis found that visually, the inner shaft was broken 0.60 inches from the distal end of the inner hub when returned.There was contamination on the inside diameters of the inner and outer distal tips.Under magnification, there were striations observed on the outside diameter of the inner shaft.The inside diameter of the outer tube shall be 0.148 ± 0.001 inches and the largest pin gage that could fit was 0.1475 inches which was in specification.The outside diameter of the inner cutter tip shall be 0.145 ± 0.001 inches and the actual measurement was between 0.1452 and 0.1453 inches which was in specification.Functionally, the device failed functional testing due to the damaged condition upon return and the inability to load the device into a handpiece.In the returned condition, there was an out of specification condition that was related to the complaint.H6: previously applied fdm b21, fdr c21, fdc d16 and img g04041 codes are no longer applicable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPS® SYSTEM NAVIGATED BLADE
Common Device Name
BUR, EAR, NOSE AND THROAT
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
justin ellis
6743 southpoint drive north
jacksonville, FL 32216
7635265677
MDR Report Key21557470
Report Number1045254-2025-00796
Device Sequence Number19941035
Product Code EQJ
UDI-Device Identifier00763000579272
UDI-Public00763000579272
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K130608
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1884380EM
Device Catalogue Number1884380EM
Device Lot Number0230022326
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/13/2025
Supplement Date Received by Manufacturer03/12/2025
Initial Report FDA Received Date03/10/2025
Supplement Report FDA Received Date04/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-