| Catalog Number |
010000999 |
| Medical Device Problem Codes |
Fracture (1260); Off-Label Use (1494)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
01/20/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that approximately two months post-implantation, the patient underwent a hip revision due to dislocation of the bearing and shell.There was a liner exchange.A screw or screw fragment was noted in the acetabulum.The head and stem are competitor products.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: cat: p0463044 lot: 0001843692 avan cmntd shell ss 44mm.Cat: 00700006220 lot: 65271449 62mm cup size revision shell.Cat: 00662406550 lot: 64825923 bone screw self-tapping.Cat: 00662406560 lot: 66840141 bone screw self-tapping.G2: foreign: australia.The customer has indicated that the product will not be returned to zimmer biomet for investigation, as it was not returned by the hospital/remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Visual examination of the provided pictures identified the explanted bearing with the head still assembled.Some bio-debris was noted.No further evaluation can be made.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Radiographs were provided and review of the available records identified the following: superolateral dislocation of the right hip arthroplasty.It was identified that zimmer biomet devices were implanted with competitor devices.Zimmer biomet has not confirmed the compatibility for these combinations of devices.It is unknown if these off-label usages may have caused or contributed to the reported events.Incompatible combinations were also identified between zb products.A definitive root cause cannot be identified.This complaint was confirmed based on the mmi review of the provided x-rays.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further information is available at the time of this report.
|
| |
|