|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
It was reported that the patient underwent a ureteroscopic balloon dilatation surgery.After the balloon dilation catheter was placed at the targeted position, observation under the microscope showed that the balloon could not remain inflated.When the balloon was observed externally, it was found that the balloon body was leaking.The balloon was replaced with a new one, and the surgery went smoothly.Patient was exposed to hazardous materials, blood or body fluids.It was unknown what medical intervention was provided.
|
|
The reported event is confirmed cause unknown.Visual evaluation noted the sample was received within a ziploc back marked with a biohazard label and sterilized by eo sterilization.No actions can be taken at this time since a root cause was not identified.Dhr review did not show any problems or conditions that would have contributed to the reported event.The instructions for use were found adequate and state the following: instructions for use: description: the x-force balloon dilation catheter is a dual lumen catheter with a balloon mounted on the distal tip.It has a radiopaque tip and two radiopaque markers beneath the balloon that define the working length.The lumen labeled with the rated burst pressure (30 atm) is for balloon inflation.The other lumen allows the catheter to track over a 0.038 (.97mm) diameter guidewire and can be used for monitoring of pressure or the infusion of medication and/or contrast medium.Each balloon inflates to a stated diameter and length at a specific pressure, typically 10 atm.The x-force balloon dilation catheter comes packaged with a refolding tool and is available with or without an inflation device.It comes sterile and is for single use only.Indications for use the x-force balloon dilation catheter is recommended for use in the dilation of the urinary tract.Contraindications: do not use the x-force balloon dilation catheter in the presence of conditions, which create unacceptable risk during the dilation of the urinary tract.Warnings: do not exceed the recommended rated burst pressure (rbp) for this device.Balloon rupture may occur if the rbp rating is exceeded.Please refer to the device label for the recommended maximum rated burst pressure.Extreme caution should be used when considering dilation of irregular, not-compliant tissue or in the presence of certain calculi.This is a single use device.Do not re-sterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Precautions: caution: federal law (u.S.A.) restricts this device to sale by or on the order of a physician.Only a physician who has an understanding of the clinical applications, technical principles and associated risks of balloon dilation within the urinary tract should use this device.After use, the x-force balloon dilation catheter and/or accessories may be considered a potential biohazard.Handle and dispose of in accordance with medical practice and applicable laws and regulations.Inspection prior to use: the x-force balloon dilation catheter is a sterile, single use device.Carefully inspect the catheter and the sterile packaging for signs of damage that may have occurred during shipment.Do not use the product if damage is evident.Preparation of the catheter all x-force balloon dilation catheters contain air in the balloon lumen.The air must be removed to allow liquid to fill the balloon when it is inflated.1.Remove the protective sheath from the balloon.2.Attach the inflation device to the connector on the balloon lumen.3.Open the stopcock and draw back on the inflation device to remove the air from the balloon catheter.4.Close the stopcock, remove the inflation device, depress the plunger to remove any air and reattach to the balloon catheter.5.Repeat steps 3-4 until all air is removed from the balloon lumen.Catheter insertion 1.Introduce the catheter carefully over a 0.038 (.97mm) guidewire and place it in the area that needs to be dilated.Use the radiopaque markers to aid in proper positioning.Note: dilation procedures should be conducted under fluoroscopic guidance with appropriate x-ray equipment or direct vision.Caution: do not advance or withdraw the catheter or guidewire against any significant resistance.The cause of the resistance must be determined fluoroscopically and remedial action taken.Inflating the balloon catheter: 1.Fill the inflation device (eagle inflation device) with sterile diluted contrast medium.2.Attach the inflation device to the balloon lumen.3.Inflate the balloon note: do not use air or any gaseous substances as a balloon inflation medium always use sterile liquid media.Note: the use of an inflation device with a pressure gauge is highly recommended to make sure adequate pressure is applied and the rated burst pressure of the balloon is not exceeded.Caution: if loss of pressure in the balloon occurs or the balloon ruptures, immediately stop the procedure, deflate the balloon and remove carefully.Do not attempt to re-inflate the balloon.If after inspection it is noted pieces of the balloon are missing, check for and endoscopically retrieve the fragments when possible.Warning: do not exceed the recommended rated burst pressure (rbp) for this device.Balloon rupture may occur if the rbp rating is exceeded.Please refer to the device label for the recommended maximum rated burst pressure.Extreme caution should be used when considering dilation of irregular, not-compliant tissue or in the presence of certain calculi.Deflating the balloon catheter deflate the balloon using an inflation device.Since deflation times vary based on balloon sizes and shaft lengths, check fluoroscopically to confirm deflation before attempting to withdraw.1.Attach the inflation device (eagle inflation device) to the balloon catheter and remove the solution from the balloon by pulling back on the inflation device.2.Remove the inflation device from the balloon catheter.This will allow ambient pressure to enter, relaxing the balloon.3.Gently withdraw the catheter.Use of a gentle counterclockwise twisting motion is recommended when removing the catheter.Caution: if resistance is felt when removing either the catheter or the guidewire from the introducer sheath, stop and consider removing them as a single unit to prevent damage to the product.Applying excessive force to the catheter can result in tip breakage or balloon separation.Using the refolding tool the balloon catheter is supplied with a refolding tool.This aids in preparing the balloon for another insertion during the procedure.To use: 1.Manually compress the balloon.2.Position the refolding tool at one end of the balloon.3.Twist the refolding tool counter clockwise and push down on the balloon until the refolding tool traverses the entire length of the balloon.4.Once the balloon is folded, remove the refolding tool and proceed with the procedure.Dispose of properly.Storage: store the balloon dilation catheter in a cool, dry, dark place.Do not store near radiation or ultraviolet light sources as these may damage product materials.Correction: d, e, h.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
It was reported that the patient underwent a ureteroscopic balloon dilatation surgery.After the balloon dilation catheter was placed at the targeted position, observation under the microscope showed that the balloon could not remain inflated.When the balloon was observed externally, it was found that the balloon body was leaking.The balloon was replaced with a new one, and the surgery went smoothly.Patient was exposed to hazardous materials, blood or body fluids.It was unknown what medical intervention was provided.
|