| Model Number |
000000000000320618 |
| Medical Device Problem Codes |
Misconnection (1399); Delivered as Unsterile Product (1421)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/13/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Consumer reported the "tear drop label" was missing from 3 pen needles before use.Stated, she still tried to use them but they would not attach to her insulin pen.Lot: 3305030.Catalog: 320618.Date of event: 2025-13-02.Samples: yes cl.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: samples were received and an investigation was performed.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed and no non-conformances were raised in association with this type of event for this lot.Embecta was able to duplicate or confirm the indicated issue and based on trend analysis no further action is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Based on the above, no additional investigation and corrective/preventative action (capa) or situational analysis (sa) is required.
|
| |
|
Search Alerts/Recalls
|