• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. NEXGEN 14MM HEIGHT SIZE F ARTICULAR SURFACE WITH HINGE POST EXTENSION; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. NEXGEN 14MM HEIGHT SIZE F ARTICULAR SURFACE WITH HINGE POST EXTENSION; PROSTHESIS, KNEE Back to Search Results
Catalog Number 00588006014
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Unspecified Infection (1930); Insufficient Information (4580)
Date of Event 02/18/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; g3; g6; h1; h2; h3; h6; h11.Visual examination of the provided intraoperative photo identified that a hinged total knee implant is in place.The femoral component appears to show signs of use and biological debris.The device history record was reviewed and no discrepancies relevant to the reported event were found.Medical records were not provided.Infection: the root cause of the reported event was determined to be unrelated to the implanted zimmer biomet device.This complaint cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
(b)(4).G2: foreign - event occurred in australia.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that the patient underwent an initial knee arthroplasty on an unknown date.Subsequently, the patient was revised due to unknown reasons.Attempts have been made, and no further information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: a1; a3; a4; b4; b5; b7; d1; d2; d4; g3; g6; h1; h2; h4; h6 an investigation of the reported event is in progress.Once the investigation is completed, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that the patient underwent a left initial knee arthroplasty on an unknown date.Subsequently, the patient was revised due to recurring infection.It was noted that the patient appeared to have femoral loosening however would have required distal femoral replacement which was unavailable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEXGEN 14MM HEIGHT SIZE F ARTICULAR SURFACE WITH HINGE POST EXTENSION
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key21559865
Report Number0001822565-2025-00564
Device Sequence Number18423766
Product Code JWH
UDI-Device Identifier00889024199439
UDI-Public(01)00889024199439(17)250630(10)64723592
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K150028
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date06/30/2025
Device Catalogue Number00588006014
Device Lot Number64723592
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/19/2025
Supplement Date Received by Manufacturer03/12/2025
05/19/2025
Initial Report FDA Received Date03/10/2025
Supplement Report FDA Received Date03/31/2025
05/22/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/06/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
Patient Weight85 KG
-
-