• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. PS TIBIAL INSERTS SZ 3, 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. PS TIBIAL INSERTS SZ 3, 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 204-23-11
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Synovitis (2094); Osteolysis (2377)
Date of Event 02/19/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The revision reported was likely the result of prosthesis wear over the course of almost 16 years of implantation and potential contributions from axial rotation mismatch between the femoral and tibial components.Potential contributions from implant alignment and/or patient-related factors to the sequence of events cannot be confirmed as the devices were not available for evaluation, radiographs were not provided, and limited clinical information was provided.The reported patellar prosthesis wear could not be confirmed as an articular surface view was not provided.D10: ps cem.Femoral size 3, left, 234-02-03, (b)(6), cem trap tib tray, 204-04-33, (b)(6).
 
Event or Problem Description
As reported, approximately 15 years and 10 months post the initial left total knee arthroplasty, the surgeon performed a synovectomy surrounding the affected knee.Mild osteolysis was present.Patella exhibited severe wear along tracking path.The tibial insert had mild wear medially and posteriorly.The femoral and tibial components were well fixed.The patella and liner were exchanged.There was no reported breakage of a device or surgical delay/prolongation.The patient was last known to be in stable condition following the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PS TIBIAL INSERTS SZ 3, 11MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key21560235
Report Number1038671-2025-01511
Device Sequence Number19941173
Product Code JWH
UDI-Device Identifier10885862048318
UDI-Public10885862048318
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K933610
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/25/2011
Device Catalogue Number204-23-11
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2025
Initial Report FDA Received Date03/10/2025
Date Device Manufactured05/26/2006
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Hospitalization;
Patient Age75 YR
Patient SexFemale
-
-