• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS, INC XCELA; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS, INC XCELA; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Model Number 44032
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bacterial Infection (1735); Fever (1858)
Date of Event 01/31/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is not available to be returned to the manufacturer for evaluation.An investigation into the root cause for the event is currently in progress.The results of the investigation will be sent via a follow up medwatch.Reference: (b)(4).
 
Event or Problem Description
On or about on (b)(6) 2021, plaintiff underwent placement of an angiodynamics vortex implantable port, model number: h787p5455k0, lot number: 5670753.The device was implanted by dr.(b)(6), m.D., at (b)(6), to facilitate chemotherapy treatment.On or about on (b)(6) 2023, plaintiff presented to (b)(6), with complaints of a fever.On or about on (b)(6) 2023, plaintiff received a call from her doctor that her blood cultures drawn on (b)(6) 2023, came back positive for a bacterial infection, and that she needs to go to the emergency room.On or about on (b)(6) 2023, plaintiff presented to the emergency room at (b)(6), for further evaluation as directed by her doctor.After evaluation, plaintiff's medical team determined that the vortex was the source of the infection and that the implant needed to be removed.On or about on (b)(6) 2023, due to her infected vortex port catheter, plaintiff underwent surgical removal of the defective device, which was removed by (b)(6), m.D., at (b)(6).
 
Additional Manufacturer Narrative
The customer's reported complaint description of patient infection cannot be confirmed given the patient centric nature of this serious adverse event (sae).No port device was returned for evaluation since there was no report of port device malfunction, damage or performance issue during use.A device history record (dhr) review of the indicated packaging lot revealed no quality related issues or manufacturing deficiencies at the time of manufacture.Sterilization load release records were reviewed for the reported packaging lot; no issues were observed.The port was implanted into the patient more than 17 months prior to the patient's infection.There was no report of port device malfunction or performance issue during use.If the manufacturing/packaging/sterilization of the port device in question was to contribute to an infection in the patient it would have likely occurred shortly after the implant date and not more than 17 months later.Device direction for use cautions that each access of the port should be performed using aseptic technique, following the institutions universal precautions.There is no indication from the reported complaint that the manufacturing/packaging/sterilization of the port device could have contributed to the patient's infection.Infection is cautioned in the device dfu as a potential complication of port system use.Labeling review: the directions for use (dfu) that is provided with the vortex port device contains the following statements: product description: ports manufactured by angiodynamics are supplied as sterile devices, and are intended for single patient use only.Indications for use: the angiodynamics port line is indicated for any patient requiring repeated access of the vascular system for delivery of medications, nutritional supplementation, fluids, blood, blood products, and sampling of blood.Dual models are indicated for combination therapy, simultaneous infusions, withdrawal of body fluids, and bolus delivery during continuous infusion.Contraindications: angiodynamics port systems should not be implanted in the presence of known or suspected infections, bacteremia, septicemia, and peritonitis, in patients who have exhibited prior intolerance to the materials of construction, or patients whose body size or tissue is insufficient to accommodate the size of the port or catheter.Potential complications: use of angiodynamics port systems involve potential risks normally associated with the insertion or use of any implanted device or indwelling catheter including but not limited to: infection, implant rejection.System maintenance: venous systems: after use, the port should be flushed with at least 20 ml of normal saline.This should be followed by a "heparin lock".If the port is not being used, a "heparin lock" should be administered every 4 weeks.Arterial systems: after use, the port should be flushed with at least 20 ml of normal saline.This should be followed by a "heparin lock".Positive pressure should be maintained on the syringe plunger at all times to minimize reflux of blood into the catheter.If port is not being used, a "heparin lock" should be administered once a week.Peritoneal systems: the intraperitoneal port should be flushed after use, or once a month if not in use, with 10 ml of heparinized saline at a concentration of 10 units/ml.General guidelines: each access of an angiodynamics port should be performed using aseptic technique.The needle should be advanced through the septum until it contacts the base of the port body.Once positioned in the septum, the needle should not be rocked or tilted.Such movement may cause septum damage.At no time should the system be left open to air.Tubing clamps should be used to prevent inadvertent air embolism.When infusing into an angiodynamics port system, do not exceed 40 psi.Do not use a syringe size smaller than 10 ml.Smaller syringes may create an overpressurized system.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XCELA
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
ANGIODYNAMICS, INC
10 glens falls technical park
glens falls NY 12801
Manufacturer (Section G)
ANGIODYNAMICS, INC
10 glens falls technical park
glens falls NY 12801
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587990
MDR Report Key21560507
Report Number1317056-2025-00059
Device Sequence Number4373561
Product Code LJT
UDI-Device IdentifierH965440320
UDI-PublicH965440320
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K153228
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2026
Device Model Number44032
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/11/2025
Supplement Date Received by Manufacturer04/03/2025
Initial Report FDA Received Date03/10/2025
Supplement Report FDA Received Date04/06/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other; Required Intervention;
Patient SexFemale
-
-