• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS, INC BIOFLO VORTEX; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS, INC BIOFLO VORTEX; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Model Number CT80LPBDVI
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 02/12/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is not available to be returned to the manufacturer for evaluation.An investigation into the root cause for the event is currently in progress.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Event or Problem Description
An end user reported that a 8f low profile plastic vortex port became dislodged and during the removal procedure, the catheter was no longer attached to the port.The patient had to have a separate ir procedure to remove the retained catheter.No further details were able to be provided.
 
Additional Manufacturer Narrative
The customer's reported complaint description of catheter detached and migrated during a port removal procedure could not be confirmed since no port/catheter complaint sample was returned to angiodynamics.Without receiving product for evaluation, we are unable to definitively determine a root cause for this incident.The port was being removed due to report that, "it had become dislodged".It is unknown if the catheter became detached from the port housing (i.E.Loose connection) or catheter fractured/detached that allowed the catheter to migrate.The port device was in situ for 47 days so loose port/catheter connection is potential root cause for why "it had become dislodged".A device history record review of the packaging/assembly/catheter lots revealed no quality related issues or manufacturing deficiencies at the time of manufacture.There have been no other reported complaints for the indicated packaging/assembly/catheter lots for similar catheter detached issues.The catheter and locking connector are provided as separate components within the port assembly kit.The end user attaches the catheter tubing to the port (and secures junction with locking connector) during the implantation procedure.Labeling review: the directions for use (dfu) that is provided in the port kit contains the following statements: absence of a blood return or a poor blood return can be a sign of a potential complication such as occlusion, kinking, breakage, pinch-off syndrome, fibrin formation, thrombosis or malposition.This should be evaluated prior to device usage.· if the patient complains of pain, or there is swelling when the device is flushed or when medication or contrast media is administered, evaluate the device for infiltration, proper needle placement, and potential complications such as occlusion, kinking, breakage, pinch-off syndrome, thrombosis or malposition.Failure to assess these complaints or observations can lead to device failure.Precautions: to avert device damage and/or patient injury during catheter placement: · carefully follow the catheter to port connection technique provided in the dfu to ensure proper device connection and to avoid catheter damage.· assure tight connection between port body and catheter.Potential complications: use of an angiodynamics port system involves potential risks normally associated with the insertion or use of any implanted device or indwelling catheter including but not limited to: - air or catheter (or catheter fragments) embolism.- catheter occlusion, malposition, dislodgement, fragmentation, migration, disconnection or rupture.Connecting the catheter: a.Place catheter lock onto catheter.The catheter lock is bi-directional and can be mounted on the catheter in either direction.B.Trim the catheter to proper length at a 90° angle allowing sufficient slack to permit body movement, port connection and verify that the catheter is not kinked.C.Advance catheter over port stem to the midway point, slightly past the barb.D.Advance catheter lock until it engages with tactile and/or audible feedback.Note: to ensure proper assembly of port and catheter lock connector, a minimal gap (less than 0.5 mm) is expected.Note: if the catheter and locking collar are connected and then disconnected, the catheter proximal end must be re-cut to ensure a safe re-connection to the port.Precaution: prior to advancing the catheter lock connector, ensure that the catheter is properly positioned.A catheter not advanced to the proper region may not seat securely and lead to dislodgement and extravasation.The catheter must be straight with no sign of kinking.A slight pull on the catheter is sufficient to straighten it.Advancing the catheter lock over a kinked catheter may damage the catheter.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOFLO VORTEX
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
ANGIODYNAMICS, INC
10 glens falls technical park
glens falls NY 12801
Manufacturer (Section G)
ANGIODYNAMICS, INC
10 glens falls technical park
glens falls NY 12801
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587990
MDR Report Key21560834
Report Number1317056-2025-00062
Device Sequence Number9215857
Product Code LJT
UDI-Device Identifier15051684016773
UDI-Public15051684016773
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K190559
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCT80LPBDVI
Device Lot Number5828230
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/12/2025
Supplement Date Received by Manufacturer07/20/2025
Initial Report FDA Received Date03/10/2025
Supplement Report FDA Received Date06/25/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/06/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-