| Model Number |
MA60AC |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A pharmacist reported that, during posterior vitrectomy surgery with insertion of intraocular lens, while injecting the lens, the doctor took the temporal haptic with the forceps to pass it through the sclera, when repeating the process with the nasal haptic, it broke in two.So, it was decided to explant the damaged lens and replace it with the another lens during the same process to complete the surgery.There was patient contact and no impact to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in d.3, d.9., h.3., h.6.And h.10.The lens was returned positioned incorrectly in the lens case.Blood and solution was dried on the lens.One haptic was broken in the distal area (not returned).The optic was cut in half, typical of insertion and removal.Product history records were reviewed and the documentation indicated the product met release criteria.The file indicated the use of a qualified cartridge and handpiece with a non-qualified viscoelastic.The reported broken haptic was observed.The root cause for the lens damage may be related to a failure to follow the instructions for use (ifu).A non-qualified viscoelastic was indicated.The ifu instructs: company foldable iols are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs to completely fill the cartridge with ovd immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.The ifu instructs: use holding forceps to grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until it is centered with or slightly past the outline etched into the top of the cartridge.The trailing haptic will extend from the proximal end of the cartridge.Position the trailing haptic to the left of the haptic post as shown in the detailed view.This will allow the plunger to go past the haptic during the initial advancement of the plunger into the cartridge.Verify the lens is positioned on the bottom surface of the cartridge.The haptic should be maintained to the left of the post when the lens is loaded correctly.Accurate positioning of the lens will decrease the potential for optic and haptic damage.The handpiece ifu instructs: important: the plunger should make initial contact with the cartridge at the ramp.In the event the plunger does not contact the cartridge at the ramp, do not use the handpiece and contact company.Verify that the plunger tip is properly engaging the lens optic and that the lens moves forward at the same rate as the plunger while slowly advancing the plunger forward to avoid damaging the lens.When the threads on the knob make contact with the barrel, turn the knob clockwise approximately half turn to engage the threads and then stop.The iol will now be in the dwell position.Inspect to ensure the plunger is behind the optic.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|