Model Number UROLIFT SYSTEM
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/13/2025
Type of Reportable Event Malfunction
Event or Problem Description
On january 13, 2025, neotract has been made aware that ¿the urolift implant misfired during procedure.¿ additional information received on january 14, 2025, confirmed that the procedure was completed successfully and that the patient was not harmed.Further additional information received on january 23, 2025, indicated that the procedure was performed on (b)(6), 2025, and the device will be returned for investigation.On february 11, 2025, the investigation of the returned sample revealed that the needle was broken in half and no pieces of the needle were missing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEOTRACT UROLIFT SYSTEM
Common Device Name
UROLIFT SYSTEM
Manufacturer (Section D)
HUDSON RESPIRATORY CARE TECATE
prolongacion mision eusebio ki
no. 1316, rancho el descanso
tecate
MX 
Manufacturer (Section G)
HUDSON RESPIRATORY CARE TECATE
prolongacion mision eusebio ki
no. 1316, rancho el descanso
tecate
MX  
Manufacturer Contact
laura gonzalez serrano
4155 hopyard road
pleasanton, CA 94588
9195448000
MDR Report Key21561902
Report Number3015181082-2025-00040
Device Sequence Number21124674
Product Code PEW
UDI-Device Identifier10814932020275
UDI-Public10814932020275
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K193269
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/29/2025
Device Model NumberUROLIFT SYSTEM
Device Catalogue NumberIPN918876
Device Lot Number73C2301015
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/11/2025
Initial Report FDA Received Date03/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/29/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-