The device was returned to stryker sustainability solutions for evaluation.Upon visual inspection of the received complaint device, evidence of clinical use was identified.The blade was fractured and separated from the device tip.The broken tip was not returned with the device.The teflon pad, handle, and contact rings appears to be intact.Further inspection reveals the blade was not centered between the actuating shaft.Appears the width of the actuating shaft was bent inward.The device was then disassembled to inspect the scalpel rod.Inspection reveals gouges could been seen on the blade from contacting the inner geometries of the shaft assembly.The fracture was found high on the scalpel rod not in the cutting area of the blade.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.The reported event could be attributed to: - applying improper or unnecessary directional or rotational force to connect instrument to hand piece - jaws/blade subassembly damage or separation - too much force or torque applied to instrument, or grasping/pulling.The instructions for use (ifu) state: - use the torque wrench (already mounted to the shaft) to tighten the blade onto the hand piece.Turn the torque wrench clockwise while holding only the gray hand piece until it clicks twice indicating that sufficient torque has been applied to secure the blade.- do not attempt to bend, sharpen, or otherwise alter the shape of the blade.Doing so may cause blade failure and user or patient injury.- avoid contact with any and all other instruments while the instrument is activated.Contact with staples, clips or other instruments while the instrument is activated may result in cracked or broken blades.- note: do not use any other means than the torque wrench to attach or detach the instrument from the hand piece.- note: do not torque the instrument by hand without the torque wrench or damage may occur to the hand piece.The reported event will continue to be monitored through post-market surveillance.
|