• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER-ENDOSCOPY LAKELAND 5300; SINGLE-USE REPROCESSED ULTRASONIC SURGICAL INSTRUMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER-ENDOSCOPY LAKELAND 5300; SINGLE-USE REPROCESSED ULTRASONIC SURGICAL INSTRUMENTS Back to Search Results
Model Number HARH45RR
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/15/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the harmonic scalpel stopped working in the middle of case.Generator and handpiece being tested as well.There was no patient injury or medical intervention and the complainant is not aware of the duration of the extended procedure time.These are commonly used devices that are readily available.
 
Additional Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Upon visual inspection of the received complaint device, evidence of clinical use was identified.The blade was fractured and separated from the device tip.The broken tip was not returned with the device.The teflon pad, handle, and contact rings appears to be intact.Further inspection reveals the blade was not centered between the actuating shaft.Appears the width of the actuating shaft was bent inward.The device was then disassembled to inspect the scalpel rod.Inspection reveals gouges could been seen on the blade from contacting the inner geometries of the shaft assembly.The fracture was found high on the scalpel rod not in the cutting area of the blade.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.The reported event could be attributed to: - applying improper or unnecessary directional or rotational force to connect instrument to hand piece - jaws/blade subassembly damage or separation - too much force or torque applied to instrument, or grasping/pulling.The instructions for use (ifu) state: - use the torque wrench (already mounted to the shaft) to tighten the blade onto the hand piece.Turn the torque wrench clockwise while holding only the gray hand piece until it clicks twice indicating that sufficient torque has been applied to secure the blade.- do not attempt to bend, sharpen, or otherwise alter the shape of the blade.Doing so may cause blade failure and user or patient injury.- avoid contact with any and all other instruments while the instrument is activated.Contact with staples, clips or other instruments while the instrument is activated may result in cracked or broken blades.- note: do not use any other means than the torque wrench to attach or detach the instrument from the hand piece.- note: do not torque the instrument by hand without the torque wrench or damage may occur to the hand piece.The reported event will continue to be monitored through post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
SINGLE-USE REPROCESSED ULTRASONIC SURGICAL INSTRUMENTS
Manufacturer (Section D)
STRYKER-ENDOSCOPY LAKELAND 5300
5300 region ct
lakeland FL 33815
Manufacturer (Section G)
STRYKER-ENDOSCOPY LAKELAND 5300
5300 region ct
lakeland FL 33815
Manufacturer Contact
andy dobos
1810 w. drake drive
tempe, AZ 85283
8888883433
MDR Report Key21561954
Report Number0001056128-2025-00012
Device Sequence Number19927281
Product Code NLQ
UDI-Device Identifier07613327346244
UDI-Public07613327346244
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K202554
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/17/2025
Device Model NumberHARH45RR
Device Catalogue NumberHARH45RR
Device Lot Number14384354
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/01/2023
Initial Report FDA Received Date03/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/17/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-