| Model Number |
SIF-Q180 |
| Medical Device Problem Code |
Residue After Decontamination (2325)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/10/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: cause traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the small intestinal videoscope exhibited foreign objects coming out of the nozzle and clogging the nozzle.There was no patient involvement.
|
| |
|
Search Alerts/Recalls
|