• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER-ENDOSCOPY PHOENIX USA; CATHETER, RECORDING, ELECTRODE, REPROCESSED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER-ENDOSCOPY PHOENIX USA; CATHETER, RECORDING, ELECTRODE, REPROCESSED Back to Search Results
Model Number 81104RR
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/15/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Upon visual inspection of the received complaint device, no relevant visible damage was detected.The catheter shaft, handle/ body, deflection and locking mechanism appeared to be intact.The device was evaluated for deflection.The catheter did not meet the curve specification.The device was evaluated using the plane template.The device met the planarity specification.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the most likely root cause is a catheter curve issue as a result of mishandling subsequent to distribution, including shipping/storage conditions or improper manipulation of the device.The instructions for use (ifu) state: - do not exert excessive pressure during placement of catheter if unknown resistance is encountered.- do not attempt to use the reprocessed ep catheter prior to completely reading and understanding the directions for use.- avoid manual pre-bending of distal curve, as this may damage steering mechanism of steerable catheters.The reported event will continue to be monitored through post-market surveillance.
 
Event or Problem Description
It was reported that the 81104 did not deflect correctly.There was no patient injury, medical intervention, or extended procedure time reported.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
CATHETER, RECORDING, ELECTRODE, REPROCESSED
Manufacturer (Section D)
STRYKER-ENDOSCOPY PHOENIX USA
10232 s. 51st st.
phoenix AZ 85044
Manufacturer (Section G)
STRYKER-ENDOSCOPY PHOENIX USA
10232 s. 51st st.
phoenix AZ 85044
Manufacturer Contact
andy dobos
1810 w. drake drive
tempe, AZ 85283
8888883433
MDR Report Key21567108
Report Number0002090040-2025-00026
Device Sequence Number17236620
Product Code NLH
UDI-Device Identifier00885825009557
UDI-Public00885825009557
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K112232
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/01/2025
Device Model Number81104RR
Device Catalogue Number81104RR
Device Lot Number5066315SH
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2023
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/27/2023
Initial Report FDA Received Date03/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-