Catalog Number 314-13-02
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Metal Related Pathology (4530)
Date of Event 02/17/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10: concomitants: (b)(6), 300-10-45 - equinoxe replicator plate 4.5mm o/s, (b)(6), 300-20-02 - equinox square torque define screw drive kit, (b)(6), 300-30-06 - equinoxe preserve stem 6mm, (b)(6), 321-20-00 - equinoxe reverse shoulder drill kit.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
 
Event or Problem Description
It was reported that the patient underwent an initial anatomic total shoulder arthroplasty on (b)(6) 2019.Subsequently, the patient underwent a revision surgery due to severe wear and metallosis on (b)(6) 2025.No further patient impact reported.No additional information available.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: d1, d4, g4, h6.Mdr section codes updated/corrected: b, c, d.The revision reported was likely the result of prosthesis fracture of the center cage leading to wear of the glenoid body and abnormal articulation between the center cage and humeral head.The cause of the fracture cannot be confirmed but may be due to off-axis drilling of the peripheral peg holes and/or due to improper initial seating of the glenoid component resulting in a rocking horse motion around a well-fixed cage.However, prosthesis wear of the humeral head and the series of events cannot be confirmed as the devices were not returned for evaluation.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE CAGE GLENOID SMALL, ALPHA
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66 court
gainesville, FL 32653
3525541928
MDR Report Key21569102
Report Number1038671-2025-01530
Device Sequence Number20919577
Product Code KWS
UDI-Device Identifier10885862172679
UDI-Public10885862172679
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K113309
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/04/2019
Device Catalogue Number314-13-02
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/17/2025
Supplement Date Received by Manufacturer10/21/2025
Initial Report FDA Received Date03/10/2025
Supplement Report FDA Received Date10/22/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/09/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-