| Model Number |
MI1210 SYNCHRONY ST |
| Medical Device Problem Code |
Circuit Failure (1089)
|
| Health Effect - Clinical Codes |
Head Injury (1879); Failure of Implant (1924)
|
| Date of Event |
03/05/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected after he hit his head.Re-implantation is being considered.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: according to the information received from the field a technical implant failure due to the reported external impact seems likely.However to determine an exact root cause device investigation would be necessary.Re-implantation is considered but no date has been scheduled yet.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected after he hit his head.Re-implantation is being considered.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected after he hit his head.The user has been reimplanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: the device investigation revealed mechanical damage to the coil pins that is typical for severe external impact to the implant site.This is a final report.
|
| |
|