| Catalog Number |
0168L16 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that patient's foley catheter was not found in patient's genitals.Nurse inflated the balloon to see if there were leaks or if the balloon had popped.The balloon was fully inflated and there were no signs or symptoms of leakage.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because no sample was returned.It is unknown whether the device had met relevant specifications.The product was used for urological care.It was unknown whether the product had caused the reported failure.The potential root cause for this failure could be user related (example: contact with sharp object)/mechanical failure/operator error/thin rubberize layer).The dhr review could not be performed without a lot number.The instructions for use were found adequate and state the following: insert foley catheters only for appropriate indications and leave in place only as long as needed.Cdc guidelines for appropriate indications for indwelling urethral catheter use: patient has acute urinary retention or bladder outlet obstruction; need for accurate urine output measurements; use for selected surgical procedures; to assist in healing of open sacral or perineal wounds; patient requires prolonged immobilization; to improve comfort for end of life care.Proper techniques for urinary catheter insertion: perform hand hygiene immediately before and after insertion insert urinary catheters using aseptic technique and sterile equipment use the smallest foley catheter possible, consistent with good drainage document the indications for catheter insertion, date and time of catheter insertion, individual who inserted catheter, and date and time of catheter removal in the patient record proper techniques for urinary catheter maintenance: secure the foley catheter.Use the statlock foley stabilization device if provided maintain a closed drainage system by utilizing pre-connected, sealed catheter-tubing junctions maintain unobstructed urine flow and keep the catheter and collection tube free from kinking keep the collection bag below the level of the bladder or hips at all times empty the collection bag regularly using a separate, clean collection container for each patient generally, drainage is accomplished by inserting the catheter through the urethra and into the bladder.However, drainage is sometimes accomplished by suprapubic or other placement of the catheter, such as nephrostomy tract.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that patient's foley catheter was not found in patient's genitals.Nurse inflated the balloon to see if there were leaks or if the balloon had popped.The balloon was fully inflated and there were no signs or symptoms of leakage.
|
| |
|
Search Alerts/Recalls
|