Catalog Number 123614A
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code Exposure to Body Fluids (1745)
Date of Event 02/21/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the foley catheter balloon was unable to deflate, and the patient was exposed to urine.It was unknown what medical intervention was provided.
 
Event or Problem Description
It was reported that the foley catheter balloon was unable to deflate, and the patient was exposed to urine.It was unknown what medical intervention was provided.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was returned.It was unknown whether the device had met relevant specifications.The potential root cause for this failure could be user related (example: over aspirated, incorrect syringe/collapse lumen/sac close eye/valve damage).A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and states the following: warning: do not use ointments or lubricants having a petroleum, they will damage latex and may cause balloon to burst.Do not aspirate urine through the drainage funnel wall.Visually inspect the product for any imperfections or surface deterioration prior to use.Use luer tip syringe to inflate with stated ml/cc od sterile water.Do not use needle or for prefilled product, remove clip and squeeze reservoir to inflate with stated ml/cc of sterile water.Recommend inflation capacities: 5ml per cc balloon: use 10ml/cc sterile water 30ml per cc balloon: use 35ml/cc sterile water.Do not exceed recommended capacities.For urological use only.To deflate catheter balloon, gently insert luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe stick in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professionals for assistance as directed by hospital protocol.Should balloon rupture occurred care should be taken to assure that all balloon fragment have been removed from the patient.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDIA® FOLEY CATHETER SILICONE COATED
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key21574478
Report Number1018233-2025-01594
Device Sequence Number18210792
Product Code KOD
UDI-Device Identifier00801741171772
UDI-Public(01)00801741171772(11)241113(17)290628(10)MYJV6172
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K922431
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number123614A
Device Lot NumberMYJV6172
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2025
Supplement Date Received by Manufacturer04/08/2025
Initial Report FDA Received Date03/11/2025
Supplement Report FDA Received Date04/23/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/13/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-