| Model Number |
A70942A |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: this event (black shadows) is caused by a component failure of the light guide bundle and the lens.This event (blurry image) is caused by a component failure of the light guide bundle and the lens.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the rigid arthroscope exhibited the light guide bundle had broken fiber and broken distal end lens.There was no patient involvement.
|
| |
|
Search Alerts/Recalls
|