| Model Number |
PED2-400-16 |
| Medical Device Problem Codes |
Device Dislodged or Dislocated (2923); Unintended Movement (3026); Migration (4003)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report that the device was successfully working after the surgery.The stent displacement was identified by the doctor a year later during a follow up evaluation.There was migration after deployment.The cause of the migration is unknown.The pipeline was implanted at the intended location.Full wall apposition was achieved.The pipeline was used for an approved (on-label) indication.The device and any accessories were prepared as indicated in the ifu.No patient symptoms or further complications were reported as a result of this event.There was no medical or surgical intervention needed to prevent a permanent impairment of a function.The event did not lead to or extend patient hospitalization.There was no injury as a result of the event.After the surgical processes, the angiographic results (xpercity evaluation) were favorable.The adverse results were identified on the follow up evaluation.The patient was undergoing surgery for treatment of an unruptured left internal carotid artery (ica) aneurysm.It was noted the patient's vessel tortuosity was normal.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Additional information received reported that the physician mentioned the event occurred after the dsa fu at 1year.They did not specified a date: dsa fu were performed between (b)(6) 2024 / (b)(6) 2025.The morphology of each aneurysm was unknown at the time of the report.B292230 - shield- hmr ¿ pipeline displaced not resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Additional information was received stating that the cause of the pipeline migration was not determined.Embolization with coils was required/planned as an additional intervention.The patient did not have any symptoms associated with the pipeline movement/migration, it was only recanalization of the aneurysm.The information is confirmed by the physician.
|
| |
|
Event or Problem Description
|
|
Additional information received clarified that reintervention was performed with coil embolization since the pipeline was no longer covering the neck of the aneurysm.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|