| Catalog Number |
30103607 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
02/19/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient underwent a revision procedure 5 months post implantation due to an anterior dislocation.Patient's soft tissue bulk had made initial approach difficult and cup version was also not ideal.There is no further information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: 110010247; item name g7 osseoti 4 hole shell 58mm g; lot # 66440575.G2: foreign: new zealand.The device will not be returned for analysis as it¿s location is not known; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: b4, g3, g6, h2, h6, and h11.Visual examination of the provided pictures identified the explanted device but could not be used to confirm the complaint.Device not returned, further analysis cannot be made.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Radiographs were provided and reviewed by a health care professional.Single undated image reviewed and not submitted to mmi.Unable to align image with event.A definitive root cause cannot be determined, however there was a competitor head used with a zimmer biomet liner.It's unknown if this caused or contributed to the reported event.As per ifu , it states 'do not use components of the g7® acetabular system with components of any other system or manufacturer, unless authorized by zimmer biomet.This complaint cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
There is no update to the prior event description provided.
|
| |
|