| Catalog Number |
SC45A |
| Medical Device Problem Codes |
Failure to Form Staple (2579); Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/26/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 03/11/25.D4: batch #unk.H4: device manufacture date is currently unavailable.A manufacturing record evaluation was performed for the finished device sc45a with lot number 182d56; and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during a lung resection procedure, it was observed that after the completion of surgical cutting, the device could not be sprung back to open the jaw.It opened after the clamp was pulled on the safety.They also noted that a few staples formed with irregular shapes, suture was used to oversew.But there is still a continuous air leakage after surgery.There was no patient consequence nor surgical delay reported.No additional information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 04/18/2025.G1: manufacturer contact name and email address corrected.An analysis of the product could not be performed since a physical sample was not received for evaluation.The lot#182d56 history records were reviewed and certified by external manufacturing that the manufacturing criteria were met prior to the release of the equipment.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product is received at a later date, the investigation will be updated as applicable.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|