| Catalog Number |
00705003400 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/20/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).G2: foreign: canada.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of black handle cup impactor has broken off into the threaded hole of the 62/64 pie impactor.There is no further information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: b4, d4: lot, d9, g3, g6, h2, h11.
|
| |
|
Event or Problem Description
|
|
There is no update to the prior event description provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4;d4;g3;h2;h3;h4;h6.Product was returned and evaluated.Visual examination of the returned product identified that the threaded end of the driver weldment was fractured; no fractured segments were returned with device.The device showed signs of use with scratches on the driver weldment shaft and guide holder.The strike plate and guide holder also had indentations.No other damage was noted during visual evaluation.Review of the device history records identified no deviations or anomalies during manufacturing.Reported event is not related to a combination of products; therefore, a compatibility review is not applicable.Medical records were not provided.A definitive root cause cannot be determined.The complaint was confirmed based on the returned, fractured device.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Event or Problem Description
|
|
This follow-up report is being submitted to relay additional information.Upon an internal data review, it was determined that this device was reported under the incorrect manufacturer.This report should be considered void, and the event will be reported under mfr 3005751028.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Upon an internal data review, it was determined that this device was reported under the incorrect manufacturer.This report should be considered void, and the event will be reported under mfr 3005751028.
|
| |
|