| Model Number |
86-0010 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: instrument has been returned, but outcome of the investigation is still pending.Review of labeling: intraoperative warnings breakage, slippage, or misuse of instruments or implant components may cause injury to the patient or operative personnel.
|
| |
|
Event or Problem Description
|
|
The surgeon reported the tip of the shoreline acs inserter broke while performing surgery on (b)(6) 2025.The tip of the inserter broke off as the surgeon was getting the cage in place.The surgeon left the cage with the inserter tip in place without the locking cover.The event did not lead to a clinically relevant increase in the duration of the surgical procedure and the surgery was completed successfully.Customer confirms there was no patient injury and the issue did not require additional medical or surgical treatment.
|
| |
|
Additional Manufacturer Narrative
|
|
Update to h3: device evaluated by manufacturer changed to "yes".Investigation conclusion: the returned inserter instrument was visually inspected to confirm the failure mode outlined in the description of the complaint.After reviewing the instrument, it was confirmed that the tip of the inner draw rod sheared off from the component.The location of the break was about 2 thread turns past where the inner draw rod is exposed from the outer sleeve locating tips.Root cause: the reported complaint may be due to misuse, surgical technique, or excessive torsional force.These devices are part of loaner sets that are subject to handling and repeat use after distribution.Wear and damage can occur over time.Exact root cause unable to be determined.
|
| |
|
Search Alerts/Recalls
|