| Catalog Number |
5-10116 |
| Medical Device Problem Code |
Unintended Deflation (4061)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed , and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
|
| |
|
Event or Problem Description
|
|
It was reported that: "i just wanted to inform you of an issue that we have been seeing with the 8.0 murphy eye tube.Per anesthesia doctors and crna's the balloon on this tube will not inflate.Thinking it was just a bad tube we opened another one and it was the same issue.There was no harm or injury to the patient.".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The customer returned four units 5-10116 et tube, cf,8.0 for investigation.Four representative samples were returned.An actual sample was not returned.The representative samples were visually examined with and without magnification.Visual examination of the returned devices revealed that the samples appear typical.No defects or anomalies were observed.A functional inspection was performed to attempt to inflate and deflate the balloon cuff on the returned et tubes.A lab inventory syringe filled with air was connected to the valve of the pilot balloon of one of the returned samples.Upon injection of air, both the pilot and cuff immediately inflated.The syringe was then removed, and the cuff and pilot balloon were inspected for leaks.No leaks were detected.The syringe was then reattached in order to deflate the balloons.Both the pilot balloon and balloon cuff were able to deflate.The sample was submerged underwater to test for any leaks.However, no leaks were detected.This process was repeated for the other samples with the same results.No issues were found with the returned samples as they were all able to properly inflate and deflate.Device history record investigation did not show issues related to complaint.The ifu states, "the tracheal tube cuff inflation system, including the valve, should be checked prior to intubation.If a malfunction is a detected in any part of the system, the tube should not be used." "care should be taken to avoid damaging the thin-walled cuffs during intubation.If cuff is damaged, the tube should not be used." the ifu also states, "cuff pressure should be monitored routinely to assure that an adequate tracheal seal is maintained and that the cuff has not become over-inflated.The tracheal tube cuff should be slowly inflated with the minimum amount of air required to provide an effective tracheal seal.Do not inflate with a measured volume of air or by "feel" of pressure in the syringe since little resistance should be felt during inflation." the reported complaint of "unable to inflate - cuff" could not be confirmed since the actual sample was not returned.The customer returned four representative samples in place of the actual sample that resulted in the complaint issue.Upon functional inspection, no problems were found with the returned devices as they were all able to properly inflate and deflate.No leaks were found with the samples.A device history record review was performed on the device with no evidence to suggest a manufacturing related cause.The probable cause of this complaint issue could not be determined without the actual device.
|
| |
|
Event or Problem Description
|
|
It was reported that: "i just wanted to inform you of an issue that we have been seeing with the 8.0 murphy eye tube.Per anesthesia doctors and crna's the balloon on this tube will not inflate.Thinking it was just a bad tube we opened another one and it was the same issue.There was no harm or injury to the patient.".
|
| |
|