| Model Number |
56300905 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/10/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H6 - neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.This regulatory report is being submitted due to retrospective review through capa 624392.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Information was received from healthcare provider (hcp) via a manufacturer representative regarding an event which occurred during a tlif (transforaminal lumbar interbody fusion) procedure.It was reported that implant would not connect tightly to inserter and would continue to rotate despite being fully-threaded.Inserter was checked for issues and there were none.He implant was partially inserted into the disc space but due to looseness, and prior to articulation, it was decided by hcp to fully remove and proceed instead with straight, static peek bullet cage, capstone ptc.There was no patient symptom reported.There were no further complications reported regarding the event.
|
| |
|
Search Alerts/Recalls
|