ZIMMER BIOMET, INC. TPRLC XR T1 PPS 14X148MM; PROSTHESIS, HIPS Back to Search Results
Model Number N/A
Medical Device Problem Code Packaging Problem (3007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/18/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).D10: 51-104130, tprlc 133 t1 pps ho 13x146mm, (b)(6), 51-107150, tprlc 133 mp type1 pps ho 15.0, (b)(6), 51-104160, tprlc 133 t1 pps ho 16x152mm, (b)(6), 51-100050, tprlc 133 fp type1 pps so 5.0, (b)(6), 51-107150, tprlc 133 mp type1 pps ho 15.0, (b)(6), 51-106170, tprlc 133 mp type1 pps so 17.0, (b)(6), 51-107150, tprlc 133 mp type1 pps ho 15.0, (b)(6), 51-109060, tloc 133 mp sp t1 ppsho 6x97.5, (b)(6), 51-101070, tprlc 133 fp type1 pps ho 7.0, (b)(6), 51-104140, tprlc 133 t1 pps ho 14x148mm, (b)(6), 51-106130, tprlc 133 mp type1 pps so 13.0, (b)(6), 51-106170, tprlc 133 mp type1 pps so 17.0, (b)(6).G2: foreign: japan.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the circulated items were inspected, and the sterile packaging was determined to be damaged.There was no patient involvement.It was reported that no further information is available.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Product was returned and evaluated.Visual evaluation of the returned product/provided photos identified damage to the sterile packaging (blister).Sterility had not been compromised.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Reported event is not related to a combination of products; therefore, a compatibility review is not applicable.The reported event did not occur in an operating room or as part of a medical procedure; medical records are not available for review.The condition of the device when it left zimmer biomet is considered conforming to specification.The root cause of the reported event can be attributed to transit damage and a packaging design issue.The reported event has been confirmed by evaluation of the returned product and provided photos.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC XR T1 PPS 14X148MM
Common Device Name
PROSTHESIS, HIPS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key21584347
Report Number0001825034-2025-00661
Device Sequence Number13125847
Product Code LPH
UDI-Device Identifier00880304517158
UDI-Public(01)00880304517158(17)320321(10)7216800
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K151448
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number51-105140
Device Lot Number7216800
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/18/2025
Supplement Date Received by Manufacturer06/06/2025
Initial Report FDA Received Date03/12/2025
Supplement Report FDA Received Date06/23/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE.
Patient SexUnknown
-
-