Blank fields on this form indicate the information is unknown or unavailable.B3: event date "first week of february" h3: device evaluation anticipated, but not yet begun.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
As reported by a distributor, a small hole was noted in a roadrunner nimble hydrophilic wire guide's sterile package.The shipping box was not damaged; however, the reporter believes that the package damage resulted from the way the devices were packaged in the shipping box.Because the hole was noted upon arrival to the distribution facility, there was no patient involvement.
|
|
Summary of event: as reported by a distributor, a small hole was noted in a roadrunner nimble hydrophilic wire guide's sterile package.The shipping box was not damaged; however, the reporter believes that the package damage resulted from the way the devices were packaged in the shipping box.Because the hole was noted upon arrival to the distribution facility, there was no patient involvement.Investigation evaluation: reviews of the complaint history, device history record (dhr), instructions for use (ifu), manufacturing instructions, and quality control procedures were conducted during the investigation.A visual inspection of the complaint device was also conducted.The complaint device was returned to cook for investigation.Indentations were noted in the clear film of the packaging, near the label.A photo provided by the customer shows holes in the package.A document-based investigation evaluation was performed.A review of the device history record found no relevant non-conformances on the lot.A review of complaint history found no additional relevant complaints for this lot number.The product ifu states ¿supplied sterilized by ethylene oxide gas in peel-open packages.Intended for one-time use.Sterile if package is unopened and undamaged.Do not use the product if there is doubt as to whether the product is sterile.Store in a dark, dry, cool place.Avoid extended exposure to light.Upon removal from package, inspect the product to ensure no damage has occurred.¿ a review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The information provided upon review of the dmr, dhr, ifu, and investigation of the returned device suggests that there is evidence the device was manufactured to specification.There is no evidence of non-conforming devices in-house or in the field.Based on the information provided and the results of the investigation, cook has concluded that shipping/handling of the device contributed to this event, as the distributor noticed the hole upon opening the package.The risk analysis for this failure mode was reviewed and no additional escalation was required.The appropriate personnel have been notified, and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|