COOK INC NCIRCLE DELTA WIRE TIPLESS STONE EXTRACTOR; FFL, DISLODGER, STONE, BASKET, URETERAL, METAL Back to Search Results
Model Number G19110
Medical Device Problem Codes Break (1069); Device Damaged by Another Device (2915)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/23/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported, during a retrograde intrarenal surgery (rirs) to remove the stone located in the left kidney, the user detected the basket wire of a ncircle delta wire tipless stone extractor was broken while capturing the stone.Its use was discontinued.The device was tested before use.The procedure was completed with another same type device.A photo of what appears to show a basket broken wire is attached.A section of the device did not remain inside the patient's body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
G4: pma/510(k) #: exempt.H3: device evaluation anticipated, but not yet begun.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Correction: h6-medical device problem code (annex a code).Investigation evaluation.Description of event: it was reported, during a retrograde intrarenal surgery (rirs) to remove the stone located in the left kidney, the user detected the basket wire of a ncircle delta wire tipless stone extractor was broken while capturing the stone.Its use was discontinued.The device was tested before use.The procedure was completed with another same type device.A photo of what appears to show a basket broken wire is attached.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Reviews of documentation including the complaint history, device history record (dhr), quality control procedures, specifications, manufacturing instructions (mi), and instructions for use (ifu), as well as a visual inspection and functional test of the returned device, were conducted during the investigation.One ncircle delta wire tipless stone extractor was returned in an open package with label.The basket wire was broken with a segment missing - not returned.Charring can be seen at break and foreign matter at basket base is preventing actuation additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record (dhr) found no non-conformances related to the reported failure mode.A review of complaint history records shows no other related complaints associated with device lot.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that the device was manufactured to specification and there is no evidence that nonconforming product exists in house or in the field.The product ifu, provides the following information to the user: precautions · the device is conductive.Avoid contact with any electrified instrument.· enclose the device in the sheath before removing from the tray/holder.· do not use excessive force to manipulate this device.Damage to the device may occur.How supplied.Upon removal from the package, inspect the product to ensure no damage has occurred.Based on the information provided, inspection of the returned device, and the results of the investigation, unintended use error caused or contributed to the event.The broken end of the wire had a charred appearance, indicating the wire broke due to exposure to a laser or other electrified device.The appropriate personnel have been notified and cook will continue to monitor for similar events.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NCIRCLE DELTA WIRE TIPLESS STONE EXTRACTOR
Common Device Name
FFL, DISLODGER, STONE, BASKET, URETERAL, METAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key21584904
Report Number1820334-2025-00249
Device Sequence Number18351020
Product Code FFL
UDI-Device Identifier00827002191105
UDI-Public(01)00827002191105(17)271021(10)16303313
Combination Product (Y/N)N
Initial Reporter CountryCN
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG19110
Device Catalogue NumberNTSED-024115-UDH
Device Lot Number16303313
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/04/2025
Supplement Date Received by Manufacturer07/28/2025
Initial Report FDA Received Date03/12/2025
Supplement Report FDA Received Date08/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/21/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age64 YR
Patient SexMale
Patient Weight75 KG
-
-